- HOME >
- For Researchers >
- Product Search >
- Search Result >
- #27788 GLP-1 (9-36/37) Assay Kit - IBL
Product Search
#27788 GLP-1 (9-36/37) Assay Kit - IBL
- Intended Use:
- Research reagents
- Measuring Method:
- ELISA
- Sample Types:
- Human, Mouse, Rat
- Measuring Samples:
- EDTA-plasma, Cell culture supernatant
- Measurement Range:
- 1.25 - 80.00 pmol/L
- Package Size1:
- 96 Well
※ The product indicated as "Research reagents" in the column Intended Use cannot be used
for diagnostic nor any medical purpose.
※ The datasheet listed on this page is sample only. Please refer to the datasheet
enclosed in the product purchased before use.
Product Overview
Product Overview
Product Code | 27788 |
---|---|
Product Name | GLP-1 (9-36/37) Assay Kit - IBL |
Intended Use | Research reagents |
Measuring Method | ELISA |
Conjugate | HRP |
Species | Human, Mouse, Rat |
Measuring Samples | EDTA-plasma, Cell culture supernatant |
Measurement Range | 1.25 - 80.00 pmol/L |
Primary Reaction | 60 minutes at 37 ℃ |
Secondary Reaction | 60 minutes at 2 - 8 ℃ |
Sensitivity | 0.36 pmol/L |
Specificity | Substance Cross-Reactivity GLP-1(9-36/37) 100% GLP-1, Active form 0.3% GLP-2 N.D. Glucagon N.D. GIP (1-42) N.D. GIP (3-42) N.D. |
Storage Condition | 2 - 8 ℃ |
Poisonous and Deleterious Substances | Not Applicable |
Cartagena | Not Applicable |
Measuring Service | Available |
Package Size 1 | 96 Well |
Remarks1 | The name of this product was changed as #27788 GLP-1 (9-36/37) Assay Kit as of 30th September 2016. The specification of this product has not been changed. The previous product name was #27788 GLP-1, Inactive form Assay Kit. |
Product Description
Product Description
GLP-1 (Glucagon Like Peptide-1) is a transcriptional product of proglucagon gene as same as glucagon and one of gastrointestinal hormone which is secreted from L cells of lower section of the small intestine. GLP-1 (7-36) amide and GLP-1 (7-37) are active form and these peptides are dissolved in blood by DDP4 (dipeptidyl peptidase-4) into GLP-1 (9-36) amide and GLP-1 (9-37) and these peptides lose its activities in short period. GLP-1 is known as incretin as same as GIP (Glucose-dependent Insulinotropic Polypeptide), it promotes secretion of insulin from β cells of the isles of Langerhans in the pancreas as glucose dependently. GLP-1 also suppresses secretion of glucagon from α cell of pancreas and concentration of glucose in blood after meal. Drug of type 2 diabetes has been developed by utilizing the functional mechanism of GLP-1 such as blood sugar control through pancreas, delayed gastric emptying and appetite suppression which contributes weight loss as extra-pancreatic effects. This ELISA kit can quantitatively measure inactive forms of GLP-1 such as GLP-1 (9-36) amide and GLP-1 (9-37) specifically. It can also quantitatively measure as total GLP-1 in EDTA-plasma which does not contains DPP4 inhibitor since active forms GLP-1 in blood are rapidly transformed to inactive forms in 2 to 5 minutes in a half period.
The control set is available for sale on request. Learn More
Keywords:
GIP / GLP-1 / Insulin / Incretin
The control set is available for sale on request. Learn More
Keywords:
GIP / GLP-1 / Insulin / Incretin
References
References
Note: Retrieve by PMID number in displayed by abstract: http://www.ncbi.nlm.nih.gov
FAQ
FAQ
-
Q.Can serum samples be used for this kit?
-
A.Not recommended to use serum samples. Please use EDTA-plasma samples.
-
Q.What is recommended anticoagulant?
-
A.We use 2Na.
-
Q.Is composition of EIA buffer of each ELISA kit all same? Can it be mixed to use?
ELISA common FAQ -
A.No it isn't. As constitute of each EIA buffer is different, it cannot be mixed with other lots or EIA buffers contained in other kind of ELISA kits.
-
Q.What is the composition of concentrated wash buffer?
ELISA common FAQ -
A.It contains ordinary Tween and phosphate buffer (0.05% Tween-20 in PB).
-
Q.What is the feature of the plate?
ELISA common FAQ -
A.We use plate that is flat bottom and removable strip type plate (8wellx 12 strips).
-
Q.Can I re-use standard after reconstitution?
ELISA common FAQ -
A.Not recommended to re-use standard after reconstitution. Please use it at once after the reconstitution.
Please note that there are some exceptions. One time freeze-thaw the standard is acceptable for use after reconstitution for some ELISAs.
Please check the details on each product datasheet. -
Q.What is different between reagent blank and test sample blank?
ELISA common FAQ -
A.Reagent blank means a well is only added EIA buffer and the purpose is confirming whether the Test sample value is influenced by lack of washing process or other operations. Test sample blank means a well is added EIA buffer and HRP antibody and the purpose is to calculate the background.
-
Q.How many samples can be measured by this kit?
ELISA common FAQ -
A.The pre-coated plate contained in our ELISA kit is 96 wells plate. We recommend to use 16 wells (2 slits) for standard and 80 wells (10 slits) for 40 samples in duplicate.
-
Q.What is LOD (Limit of Detection)?
ELISA common FAQ -
A.It (LOD) is defined as sensitivity that is calculated using the NCCSL method. Please refer to a datasheet of each product.
-
Q.What is LOQ (Limit of Quantification)?
ELISA common FAQ -
A.It (LOQ) is the lowest value of measurement (standard) range. Please refer to a datasheet of each product.
-
Q.What is the definition of Over Night (O/N) reaction?
ELISA common FAQ -
A.It means that the reaction is required more than 16 hours unless otherwise specifically defined it on a datasheet of each ELISA product.
-
Q.What is the specification of quality control for ELISA product release?
ELISA common FAQ -
A.The information of specification is available on individual lot specific CoA. Please contact us with your reference lot number for obtaining of specific CoA.
-
Q.What is the number (e.g. 432143214321) at the edge of strips of the plate?
ELISA common FAQ -
A.According to the plate maker (ThermoFisher), it does not have any specific meaning as it is just the number of molds.
-
Q.How to wash an ELISA plate?
ELISA common FAQ -
A.Washing it by an auto-washer is highly recommended.
If it is not available, please refer to the demo video (only 2 mins) using a washing bottle. -
Q.The wells turned black during the test with the kit.
ELISA common FAQ -
A.It is possible that the wells were not washed sufficiently during the washing process after the HRP-labeled antibody reaction.
Be sure to wash the wells enough times as described in the data sheet with washing buffer of more than 350 µL. -
Q.How do you define the acceptance range of quality control of control sets?
Control Set Common FAQ -
A.In principal, it is defined as ±15%-20% based the standard value described on each datasheet.
-
Q.What is the shelf life of control set?
Control Set Common FAQ -
A.The shelf life is 12 months upon the goods is dispatched from our lab if the goods is unopned.
-
Q.What is the storage condition of control sets?
Control Set Common FAQ -
A.It should be stored under chilled condition (2~10℃).
-
Q.Can the control set be used after reconstitution?
Control Set Common FAQ -
A.No it cannot be re-used after the reconstitution.
-
Q.What is the package form of control sets?
Control Set Common FAQ -
A.It is lyophilized form.